Sensor Alignment and Calibration
IM-Mercury_II_6000 Series Rev. 1 Page 14 ©2014 MicroE Systems
Mercury II 6000 Series Encoders
Installation Manual and Reference Guide
5.2 Using the Software
Use the following instructions to perform sensor alignment and calibration using the Alignment
Tool and the SmartPrecision II Software.
5.2.1 Connect the Alignment Tool and Encoder
Verify that the Alignment Tool Adapter/SmartPrecision II Alignment Tool is not powered on. The
Power/Cal. LED should be off.
Connect the MII 6000 encoder and Ethernet cable as shown below.
Power up the Alignment Tool by plugging in 5VDC to the Alignment Tool Adapter.
Result: The Power/Cal. LED illuminates a steady green. The sensor is initialized after an
approximate 10 second delay.
Computer
(Windows or
other OS)